Gelişmiş Arama

Basit öğe kaydını göster

dc.contributor.authorYildiz, Ergul
dc.contributor.authorSaracoglu, Kemal Tolga
dc.contributor.authorSaracoglu, Ayten
dc.contributor.authorSorbello, Massimiliano
dc.contributor.authorKizilay, Deniz
dc.contributor.authorKafali, Haluk
dc.date.accessioned2020-12-02T18:01:54Z
dc.date.available2020-12-02T18:01:54Z
dc.date.issued2019
dc.identifier.issn1642-5758
dc.identifier.issn1731-2531
dc.identifier.urihttps://doi.org/10.5114/ait.2019.91193
dc.identifier.urihttp://hdl.handle.net/11446/3760
dc.descriptionWOS: 000505051800006en_US
dc.descriptionPubMed: 31893603en_US
dc.description.abstractBackground: Guidelines for management of unanticipated difficult intubation recommend the use of supraglottic airway devices (SADs) in cases of failed intubation. However, there is a lack of comparative studies for different type of devices. in this randomised controlled trial, the performance of 1st and 2nd generation supraglottic airway devices was compared in patients with a simulated difficult airway. Material and methods: We enrolled 90 patients, scheduled for elective surgery and suitable for supraglottic airway device insertion. Laryngeal mask airway (LMA)-classic (LMAC), LMA-proseal (LMAP) and LMA-flexible (LMAF) were evaluated. the modified Mallampati test was used for the preoperative airway assessment. Maximal mouth opening, body mass index, thyromental and sternomental distances, and neck circumference were measured, and patients with predicted difficulty were excluded. Insertion time, ease of insertion, oropharyngeal leak pressure, and Brimacombe and Berry Bronchoscopy Scores were evaluated. Peak airway pressure was measured at 1, 15 and 60 min following the insertion of SADs. Complications were recorded. Results: Oropharyngeal leak pressures were 35.2 +/- 8.1, 31.7 +/- 7.7 and 31.3 +/- 6.0 mm Hg for LMAP, LMAC and LMAF respectively (P = 0.079). First min peak airway pressure values were 14.0 +/- 4.2, 15.0 +/- 3.9, 14.9 +/- 4.4 mm Hg respectively (P = 0.403). There was a significant positive correlation between oropharyngeal leak pressure and first min peak airway pressure (r = 0.264, P = 0.013). Mean number of attempts was 1.1 +/- 0.3 times (P = 0.840). Insertion time was 20.0 +/- 10.4, 17.0 +/- 5.7 and 16.4 +/- 10.2 s respectively (P = 0.440). Ease of insertion score was 2.0 +/- 0.9, 2.1 +/- 0.9 and 2.1 +/- 1.3 respectively (P = 0.837). There was no significant difference for optimization manoeuvre requirement or fibreoptic scope grades (P = 0.265, P = 0.651, respectively). Conclusions: First and second generation of supraglottic airway devices provided similar clinical performance for patients with difficult airway and trauma due to limited cervical motion.en_US
dc.language.isoengen_US
dc.publisherTermedia Publishing House Ltden_US
dc.identifier.doi10.5114/ait.2019.91193en_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectdifficult airwayen_US
dc.subjectcervical collaren_US
dc.subjectsupraglottic airway devicesen_US
dc.titlePerformance of first and second generation supraglottic airway devices in patients with simulated difficult airway: a randomised controlled trialen_US
dc.typearticleen_US
dc.relation.journalAnaesthesiology Intensive Therapyen_US
dc.departmentDBÜen_US
dc.identifier.issue5en_US
dc.identifier.volume51en_US
dc.identifier.startpage373en_US
dc.identifier.endpage379en_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.department-temp[Yildiz, Ergul; Saracoglu, Kemal Tolga; Saracoglu, Ayten; Kizilay, Deniz; Kafali, Haluk] Istanbul Bilim Univ, Dept Anaesthesiol & Intens Care, Istanbul, Turkey; [Sorbello, Massimiliano] AOU Policlin Vittorio Emanuele Catania, Dept Anesthesiol & Intens Care, Catania, Italyen_US


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster